MRI with a pacemaker or other active implant: what to check
Pacemakers, ICDs, neurostimulators and other devices with electronics or batteries: what MR Conditional means and what has to happen before the scan.
Active implants
Devices with electronics, batteries or leads can malfunction, heat or deliver unintended stimulation.
- Pacemakers and ICDs: an MR Conditional system means generator and leads together. Follow the stated conditions; the device is programmed to its MR mode before the scan and restored after, with ECG and pulse oximetry monitoring during it
- Non-conditional or mixed cardiac systems, abandoned or fractured leads: only under a written institutional protocol with cardiology and the MR medical director
- Deep brain, spinal cord and vagus nerve stimulators: conditions are strict (often a low SAR or B1+rms limit, a transmit-receive head coil only, impedance check, stimulation off)
- Cochlear implants: the internal magnet can move or demagnetize. Many need a head bandage or magnet removal at 1.5 T; check the exact model
- Insulin pumps and glucose sensors: remove before Zone IV
- Programmable shunt valves: the setting can change in the field; have it checked after the scan
- Implanted drug pumps: the motor can stall during the scan; follow the manufacturer's check afterwards
Lead-tip heating can injure the heart or brain; a reprogrammed device can stop pacing.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.
References
- ACR Manual on MR Safety, 2024 edition, with the changes effective February 2026. American College of Radiology.
- ACR Manual on Contrast Media, including the 2023 gadopiclenol and 2024 classification updates. ACR Committee on Drugs and Contrast Media.
- Weinreb JC et al. Use of intravenous gadolinium-based contrast media in patients with kidney disease: ACR–NKF consensus statements. Radiology 2021;298:28–35.
- IEC 60601-2-33, Edition 4.0 (2022). Steckner MC et al. Transitioning from 0.5 to 0.9 mT. Magn Reson Med 2024;92:2237–2245.
- ASTM F2503. Standard practice for marking medical devices and other items for safety in the MR environment.
- Turkbey B et al. PI-RADS version 2.1. Eur Urol 2019;76:340–351.
- ACR LI-RADS v2018 CT/MRI core and technical recommendations.
- Mann RM et al. Breast MRI: guidelines from the European Society of Breast Imaging. Eur Radiol 2008;18:1307–1318. ACR Breast MRI Accreditation Program requirements.
- Kramer CM et al. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson 2020;22:17.
- Wattjes MP et al. 2021 MAGNIMS–CMSC–NAIMS consensus recommendations on the use of MRI in patients with multiple sclerosis. Lancet Neurol 2021;20:653–670.
- Bernasconi A et al. Recommendations for the use of structural MRI in the care of patients with epilepsy (HARNESS-MRI). Epilepsia 2019;60:1054–1068.
- Beets-Tan RGH et al. MRI for clinical management of rectal cancer: updated ESGAR consensus recommendations. Eur Radiol 2018;28:1465–1475.
- Taylor SA et al. The first joint ESGAR/ESPR consensus statement on the technical performance of cross-sectional small bowel and colonic imaging. Eur Radiol 2017;27:2570–2582.
- Lambert RGW et al. Standardised image acquisition protocol for MRI of the sacroiliac joints: ASAS–SPARTAN consensus. Ann Rheum Dis 2024;83:1628–1635.
- Radiopaedia.org: MRI protocol, sequence and artifact articles. Elster AD: MRIquestions.com (Questions and Answers in MRI).
- Siemens Healthineers: MRI Acronyms, Cross-Vendor Comparisons. Operator manuals and application guides: GE HealthCare, Philips, Siemens Healthineers.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.