MR safety and patient preparation

MRI with metal in the body: clips, implants and foreign bodies

Aneurysm clips, metal fragments in the eye, joint replacements, stents and shrapnel: when the scan can go ahead and when it cannot.

Passive implants and foreign bodies

Purpose

Metal without electronics can still move, heat, or hide anatomy.

Key points
  • Intracranial aneurysm clips: scan only with written documentation of the exact make and model showing it is safe at your field strength
  • Metal fragment in or near the eye: do not scan until excluded
  • Orthopaedic plates, screws and joint replacements: firmly fixed and not a displacement risk; the issues are heating over long hardware and image artifact
  • Coronary stents, heart valve prostheses, sternal wires: almost all are MR Conditional; check the conditions for the model
  • Vascular coils, filters and weakly ferromagnetic devices: some need a waiting period after implantation; follow the label
  • Breast tissue expanders with a magnetic port, some older ear implants, magnetic dental or eyelid implants: usually MR Unsafe
  • Bullets and shrapnel: the decision depends on composition and distance from vital structures; refer to the radiologist
  • Remove piercings, jewellery, hair accessories, removable dental work and hearing aids
  • Intrauterine devices: copper types are MR Conditional; check stainless-steel types
If ignored

Deaths have followed scans of patients with ferromagnetic aneurysm clips that were assumed to be safe.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.

References

  • ACR Manual on MR Safety, 2024 edition, with the changes effective February 2026. American College of Radiology.
  • ACR Manual on Contrast Media, including the 2023 gadopiclenol and 2024 classification updates. ACR Committee on Drugs and Contrast Media.
  • Weinreb JC et al. Use of intravenous gadolinium-based contrast media in patients with kidney disease: ACR–NKF consensus statements. Radiology 2021;298:28–35.
  • IEC 60601-2-33, Edition 4.0 (2022). Steckner MC et al. Transitioning from 0.5 to 0.9 mT. Magn Reson Med 2024;92:2237–2245.
  • ASTM F2503. Standard practice for marking medical devices and other items for safety in the MR environment.
  • Turkbey B et al. PI-RADS version 2.1. Eur Urol 2019;76:340–351.
  • ACR LI-RADS v2018 CT/MRI core and technical recommendations.
  • Mann RM et al. Breast MRI: guidelines from the European Society of Breast Imaging. Eur Radiol 2008;18:1307–1318. ACR Breast MRI Accreditation Program requirements.
  • Kramer CM et al. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson 2020;22:17.
  • Wattjes MP et al. 2021 MAGNIMS–CMSC–NAIMS consensus recommendations on the use of MRI in patients with multiple sclerosis. Lancet Neurol 2021;20:653–670.
  • Bernasconi A et al. Recommendations for the use of structural MRI in the care of patients with epilepsy (HARNESS-MRI). Epilepsia 2019;60:1054–1068.
  • Beets-Tan RGH et al. MRI for clinical management of rectal cancer: updated ESGAR consensus recommendations. Eur Radiol 2018;28:1465–1475.
  • Taylor SA et al. The first joint ESGAR/ESPR consensus statement on the technical performance of cross-sectional small bowel and colonic imaging. Eur Radiol 2017;27:2570–2582.
  • Lambert RGW et al. Standardised image acquisition protocol for MRI of the sacroiliac joints: ASAS–SPARTAN consensus. Ann Rheum Dis 2024;83:1628–1635.
  • Radiopaedia.org: MRI protocol, sequence and artifact articles. Elster AD: MRIquestions.com (Questions and Answers in MRI).
  • Siemens Healthineers: MRI Acronyms, Cross-Vendor Comparisons. Operator manuals and application guides: GE HealthCare, Philips, Siemens Healthineers.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.