Artifacts & problems

What each artifact looks like, why it happens and how to fix it, plus common scanner problems. 20 free pages.

Artifacts: cause and fix

Magic angle artifact in MRI: how to recognise it and confirm it

A tendon that looks bright on a short-TE image is not always torn. The magic angle effect is the common mimic.

Aliasing (wrap-around) in MRI: cause and fixes

Anatomy outside the field of view folds back into the image. These are the fixes and what each one costs.

Pulsation and flow ghosts in MRI: cause and fixes

Copies of a vessel or of CSF appear at regular intervals along the phase direction and can look like lesions. This is how to recognise them and move…

Chemical shift artifact in MRI: the dark and bright bands at fat-water borders

Fat and water are placed a few pixels apart along the frequency direction. Bandwidth and field strength decide how far.

India ink artifact: the black outline on opposed-phase images

A thin black line around organs on opposed-phase gradient echo is expected, and it is useful. This is where it comes from.

Metal and susceptibility artifact in MRI: how to reduce it

Signal voids, bright pile-up and distortion near metal or air. These are the changes that help most, and what each one costs.

Gibbs ringing (truncation artifact) in MRI

Fine parallel lines next to a sharp edge, such as the cord or the skull. They come from sampling too few lines of k-space.

Zipper artifact and RF interference in MRI

A dotted line across the image usually means outside RF is getting into the room. Check these things before calling service.

Partial volume effect in MRI

A voxel that holds two tissues shows the average of both. Small structures blur or disappear.

Cross-talk between MRI slices: cause and fixes

Slices placed too close together excite each other and lose signal. The slice gap and the acquisition order fix it.

Dielectric shading at 3 T: the dark patch in the abdomen

A dark or bright area in the middle of a large body part at 3 T is often the RF field and not disease.

Parallel imaging artifacts in MRI (ASSET, SENSE, GRAPPA)

Noise bands or a ghost in the centre of the image after turning acceleration up. This is what went wrong.

EPI distortion and the N/2 ghost on DWI

Diffusion images stretch near air and can show a half-FOV ghost. These are the causes and the fixes.

Spike (herringbone) artifact in MRI

A criss-cross pattern over the whole image comes from one bad data point. Find out whether it is the room or the scanner.

Inflow and CSF flow effects in MRI

Bright vessels on the first slices and dark or bright CSF that looks like a lesion. Flow explains both.

Banding artifact on balanced SSFP (FIESTA, bFFE, TrueFISP)

Dark curved bands on FIESTA-type images follow field inhomogeneity. Shim and TR move them out of the anatomy.

Coil shading and uneven signal in MRI

One side of the image is brighter than the other. Check the coil and the intensity correction before repeating the scan.

Scanner problems before you call the engineer