Knee MRI planning and protocol: slice positioning and sequences

How to plan a knee MRI: slice planning drawings for the sagittal, coronal and axial planes, plus one common knee protocol with what each sequence is there to show.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Knee

Indications

Meniscal tear, cruciate and collateral ligament injury, cartilage damage, bone bruise.

Set-up

Knee coil, 10–15° of flexion and slight external rotation. FOV 14–16 cm, 3 mm slices.

Slice planning
Sagittal Perpendicular to the line joining the posterior femoral condyles, from one collateral ligament to the other.
Coronal Parallel to the line joining the posterior femoral condyles, from the patella to behind the condyles.
Axial From above the patella to the tibial tuberosity, perpendicular to the long axis of the leg.

Images: KMAR-50K open dataset, Wang & Shi, Mendeley Data (CC BY 4.0). Planning lines drawn by MRI Technologist Reference.

Sequences and why
Sag PD fat-satMenisci, cartilage, marrow edema, cruciate ligaments.
Sag PD or T1 without fat satMeniscal tears are best seen on short TE; marrow anatomy.
Cor PD fat-satCollateral ligaments, meniscal body and roots.
Cor T1Fracture line, marrow.
Ax PD fat-satPatellofemoral cartilage, retinacula.
Practical points
  • Popliteal artery ghosts: phase direction superior-inferior on sagittal and right-left on axial keeps them off the joint
  • Slight external rotation puts the ACL in the sagittal plane
  • Keep TE below about 25 ms on the non-fat-sat PD or meniscal tears lose contrast

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.